Reconstruction of the parameters of a soft X-radiation spectrum from signals of vacuum X-ray diodes

Citation
Av. Branitskii et Gm. Oleinik, Reconstruction of the parameters of a soft X-radiation spectrum from signals of vacuum X-ray diodes, INSTR EXP R, 43(4), 2000, pp. 486-492
Citations number
17
Categorie Soggetti
Instrumentation & Measurement
Journal title
INSTRUMENTS AND EXPERIMENTAL TECHNIQUES
ISSN journal
00204412 → ACNP
Volume
43
Issue
4
Year of publication
2000
Pages
486 - 492
Database
ISI
SICI code
0020-4412(200007/08)43:4<486:ROTPOA>2.0.ZU;2-P
Abstract
A multichannel measuring complex based on vacuum X-roy diodes with various tilters and a signal processing technique enabling to obtain a radiation po wer with a nanosecond time resolution in several spectral ranges from 0.1 t o 4 keV, power in the full spectral range, and average quantum energy are d escribed. This technique was applied on the Angara-5-1 setup in experiments on implosion of cascade liners and with a z-pinch.