Numerous methods are proposed to quantify antithyroid peroxidase autoantibo
dies. No standardization exists but most assays use the standard MRC 66/387
with a calibration factor. Costs of the tests vary between the different k
its. We evaluated the concordance of eight peroxidase autoantibodies assay
kits in two centres, using a panel of sera from 269 subjects: controls (n =
100), patients with autoimmune thyroid disease (n = 77; Graves' disease, H
ashimoto's thyroiditis), patients with non-autoimmune thyroid disease (n =
69; nodular goiter, differentiated thyroid carcinoma) and individual sera w
ith thyroglobulin antibodies only (n = 23). The concordance between the eig
ht methods was high, ranging from 88.3% to 98.8% with the total panel of se
ra. The majority of assays demonstrated high diagnostic performance. We enc
ountered some false-positive results at borderline positive levels, and the
nonrecognition of some sera by competitive assays.