Mjg. De La Huebra et al., Determination of carbendazim in soil samples by anodic stripping voltammetry using a carbon fiber ultramicroelectrode, FRESEN J AN, 367(5), 2000, pp. 474-478
A method for the determination of carbendazim (MBC) by anodic stripping vol
tammetry using a carbon fiber ultramicroelectrode was developed. The ultram
icroelectrode was made in our laboratory and its electrochemical behavior w
as characterized by measuring the electrochemical response with a solution
of potassium ferricyanide. The optimum parameters used for the determinatio
n of MBC are the following: 0.05 M phosphate buffer at pH 2.0 as supporting
electrolyte; a scan rate of v = 10.00 V s(-1) and an accumulation potentia
l of E-ac = 0.00 V. The MBC was determined in a soil sample with the method
proposed and the results found were comparable to those obtained by HPLC.