The association by aligned direct bonding of an evanescent wave optical sen
sor and small-volume flow cells for visible absorption detection is demonst
rated. We show that for sample volumes ranging from 760 to 190 nr and a sen
sor interaction length of I mm, very small cross-sectional dimensions of fl
uid channels are no more a serious obstacle for optical detection. The lowe
st limit for the channel depth Is the penetration depth of the evanescent w
ave. which is usually in the order of magnitude of a few micrometers. (C) 2
000 Elsevier Science S.A. All rights reserved.