H. Nagahashi et al., Improved sensitivity in the measurement of residual leukocytes in plateletproducts using an automated leukocyte counter, VOX SANGUIN, 79(1), 2000, pp. 34-39
Background and Objectives: Leukocytes remaining in platelet concentrates ma
y be responsible for side effects caused by platelet transfusion. A simple
method of high sensitivity for determining trace numbers of leukocytes is c
urrently needed.
Materials and Methods: An automated leukocyte counter, the LD-1000, the pri
nciple of which is a combination of spinning down on the observation field
of stained nuclei derived from a 100 mu l neat sample and image digitizatio
n with a charge coupled device camera, was newly developed and tested for s
ensitivity and reproducibility,
Results: While the theoretical lower limit of detection was 0.01 cells/mu l
, the limit of detection of the LD-1000 was verified to be 0.2 cells/mu l.
A good correlation (r = 0.86) was observed between the results obtained usi
ng the Nageotte method and the LD-1000, Repeated measurements also confirme
d satisfactory reproducibility.
Conclusions: The instrument provides a new method of enumerating residual l
eukocytes in platelet products with better sensitivity and easier procedure
compared to the Nageotte method and will be useful for quality assurance p
urposes at blood centers. Copyright (C) 2000 S. Karger AG, Basel.