ANALYTICAL AND CLINICAL-EVALUATION OF THE IMMULITE(R) ESTRADIOL ASSAYIN SERUM FROM PATIENTS UNDERGOING IN-VITRO FERTILIZATION - ESTRADIOL INCREASE IN MATURE FOLLICLES
F. Dancoine et al., ANALYTICAL AND CLINICAL-EVALUATION OF THE IMMULITE(R) ESTRADIOL ASSAYIN SERUM FROM PATIENTS UNDERGOING IN-VITRO FERTILIZATION - ESTRADIOL INCREASE IN MATURE FOLLICLES, Clinical chemistry, 43(7), 1997, pp. 1165-1171
We examined an immunoassay for estradiol (E2) on the Immulite(R)-an au
tomated, random access chemiluminescent immunoassay system-to determin
e its accuracy and precision required for in vitro fertilization (IVF)
studies. The assay, which has a reportable range from 73 to 7300 pmol
/L, demonstrated good linearity under dilution, a detection limit of 4
4 pmol/L, and interassay CVs of 12.6% and 7.6% at 466 and 6164 pmol/L,
respectively. In a retrospective analysis of 545 serum samples, the a
ssay showed adequate agreement with an antibody-coated-tube RIA. The t
wo E2 assays showed good agreement, even on samples from patients rece
iving a variety of different estrogen replacement therapies. Longitudi
nal studies of individual IVF cycles showed good parallelism between t
he automated system and the RIA, and results by the automated assay co
rrelated well with the total number of follicles.