Ka. Snyder et al., Using impedance spectroscopy to assess the viability of the rapid chloridetest for determining concrete conductivity, J RES NAT I, 105(4), 2000, pp. 497-509
The suitability of using the initial current from the rapid chloride test (
ASTM C 1202) to determine specimen conductivity is tested using impedance s
pectroscopy with a frequency spectrum of 10 Hz to 1 MHz. The specimen condu
ctivity has an analytical relationship to specimen diffusivity and so is a
useful quantity in service life prediction. Measurements made on specimens
of different lengths indicate that the total charge passed during the six h
our conduction test carried out according to ASTM C 1202 is not a direct me
asure of specimen conductivity. Further, ohmic heating during the 6 hour te
st makes it nearly impossible to directly measure any specimen transport pr
operty from the results. The total charge passed during the 6 hour conducti
on test is, therefore, not a reliable quantity for service life prediction.
Results indicate that the direct current (dc) measurement of resistance us
ing a voltage of 60 V is sufficient to overwhelm polarization effects, ther
eby yielding an accurate estimate of the true specimen conductivity. Impeda
nce spectroscopy measurements also indicate that corrosion may form on the
brass electrodes, adding bias to a conductivity estimate based upon a dc me
asurement.