Sl. Pinski, Inappropriate pacing due to autoperpetuation of the ventricular rate stabilization algorithm: A manifestation of T wave oversensing by ICDs, PACE, 23(9), 2000, pp. 1446-1447
Citations number
9
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
Preventive pacing algorithms designed to eliminate the pause that follows a
premature ventricular depolarization have been incorporated in current imp
lantable defibrillators. We report a patient in whom intermittent T wave ov
ersensing frequently invoked the ventricular rate stabilization algorithm a
nd resulted in periods of inappropriate VVI pacing. The problem was solved
by decreasing the maximum sensitivity from 0.3 mV to 0.45 mV. Implant testi
ng had revealed adequate sensing of ventricular fibrillation with a maximum
sensitivity of 1.2 mV.