Inappropriate pacing due to autoperpetuation of the ventricular rate stabilization algorithm: A manifestation of T wave oversensing by ICDs

Authors
Citation
Sl. Pinski, Inappropriate pacing due to autoperpetuation of the ventricular rate stabilization algorithm: A manifestation of T wave oversensing by ICDs, PACE, 23(9), 2000, pp. 1446-1447
Citations number
9
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
Journal title
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
ISSN journal
01478389 → ACNP
Volume
23
Issue
9
Year of publication
2000
Pages
1446 - 1447
Database
ISI
SICI code
0147-8389(200009)23:9<1446:IPDTAO>2.0.ZU;2-8
Abstract
Preventive pacing algorithms designed to eliminate the pause that follows a premature ventricular depolarization have been incorporated in current imp lantable defibrillators. We report a patient in whom intermittent T wave ov ersensing frequently invoked the ventricular rate stabilization algorithm a nd resulted in periods of inappropriate VVI pacing. The problem was solved by decreasing the maximum sensitivity from 0.3 mV to 0.45 mV. Implant testi ng had revealed adequate sensing of ventricular fibrillation with a maximum sensitivity of 1.2 mV.