E. Abad et al., Evaluation of a new automated cleanup system for the analysis of polychlorinated dibenzo-p-dioxins and dibenzofurans in environmental samples, J CHROMAT A, 893(2), 2000, pp. 383-391
An automated cleanup system (Power-Prep(TM), Fluid Management Systems, Inc.
) was evaluated for the analysis of polychlorinated dibenzo-p-dioxins and d
ibenzofurans (PCDD/PCDF) in environmental samples. The study was firstly ca
rried out against standard mixtures containing labeled and unlabeled PCDD/P
CDF and a certified reference material (fly ash CRM 490). Next, a compariso
n between a conventional cleanup procedure and the new automated system for
PCDD/PCDF analysis in environmental samples such as flue gas emissions, fl
y ashes, sludges, ambient air and soils was also made. The suitability of t
he system was finally assessed from the participation in an interlaboratory
exercise. Owing to its capability to successfully perform the sample clean
up the Power-Prep(TM) apparatus is an alternative to the conventional metho
ds. (C) 2000 Elsevier Science B.V. All rights reserved.