Measurement of 15 MeV gamma-rays with the Ge cluster detectors of EUROBALL

Citation
B. Million et al., Measurement of 15 MeV gamma-rays with the Ge cluster detectors of EUROBALL, NUCL INST A, 452(3), 2000, pp. 422-430
Citations number
13
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
452
Issue
3
Year of publication
2000
Pages
422 - 430
Database
ISI
SICI code
0168-9002(20001001)452:3<422:MO1MGW>2.0.ZU;2-2
Abstract
A measurement of the response to 15.1 MeV gamma-rays has been made for the Ge cluster detectors in the EUROBALL array. Each cluster detector consists of seven germanium capsules surrounded by a single anticompton shield of EG O. The reaction D(B-11,gamma)C-12 + n at E-beam = 19.1 MeV has been employe d. The "adding-back" of signals simultaneously present in the capsules comp osing each cluster detector has been made on an event by event basis. The i ntensity in full-energy peak increases by a factor of three as compared to that of the spectrum obtained by summing the individual spectra of the 7 ca psules. The pulse height to energy conversion is found to be very linear fr om few hundreds keV to 15 MeV. The efficiency is discussed relative to that of large volume BaF2 scintillators. (C) 2000 Elsevier Science B.V. All rig hts reserved.