Nickel(II) ion-selective electrode based on 2-methyl-4-(4-methoxy phenyl)-2,6-diphenyl-2H-thiopyran

Citation
Mr. Ganjali et al., Nickel(II) ion-selective electrode based on 2-methyl-4-(4-methoxy phenyl)-2,6-diphenyl-2H-thiopyran, ELECTROANAL, 12(14), 2000, pp. 1138-1142
Citations number
42
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ELECTROANALYSIS
ISSN journal
10400397 → ACNP
Volume
12
Issue
14
Year of publication
2000
Pages
1138 - 1142
Database
ISI
SICI code
1040-0397(200010)12:14<1138:NIEBO2>2.0.ZU;2-A
Abstract
A new PVC membrane electrode for nickel ion based on a recently synthesized 2-methyl-4-(4-methoxy phenyl)-2,6-diphenyl-2H-thiopyran as membrane carrie r was prepared. The sensor exhibits a Nernstian response for Ni2+ over a wi de concentration range (5 x 10(-2) 2 x 10(-5) M) with a limit of detection of 9 x 10(-6) M. The response time of the sensor is ca. 10 s and the membra ne can be used for more than 8 weeks without observing any deviation. The p roposed membrane sensor shows excellent discriminating ability towards Ni2 ion in comparison to several alkali, alkaline earth, transition and heavy metal ions. It was used as an indicator electrode in potentiometric titrati on of Ni ions in bath water and 85% acetonitrile solutions.