This paper reports two unusual instances of ethylene oxide burns that were
caused by the blood pressure cuff sterilized with ethylene oxide. It is sug
gested that the mechanical compression caused by the blood pressure cuff fa
cilitated penetration of ethylene oxide residues into tissues and contribut
ed to the degree of tissue damage. (C) 2000 Elsevier Science Ltd and ISBI.