Feedback and diagnostic signals in direct sample insertion inductively coupled plasma-atomic emission spectrometry (DSI-ICP-AES)

Citation
Tj. Wood et V. Karanassios, Feedback and diagnostic signals in direct sample insertion inductively coupled plasma-atomic emission spectrometry (DSI-ICP-AES), CAN J A S S, 45(2), 2000, pp. 48-52
Citations number
21
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
CANADIAN JOURNAL OF ANALYTICAL SCIENCES AND SPECTROSCOPY
ISSN journal
12056685 → ACNP
Volume
45
Issue
2
Year of publication
2000
Pages
48 - 52
Database
ISI
SICI code
1205-6685(2000)45:2<48:FADSID>2.0.ZU;2-F
Abstract
Diagnostic signals are discussed in conjunction with an automated direct sa mple insertion-inductively coupled plasma-atomic emission spectrometry (DSI -ICP-AES) system and a computer-controlled feedback loop. Such signals were obtained from computer-controlled mass flow controllers and by monitoring emission from Ar and C spectral lines in real-time. Use of feedback and dia gnostic signals proved to be a particularly attractive way of enhancing the operational reliability of the automated DSI-ICP-AES system used for this work and at improving the overall quality of analytical data obtained with this system.