P. Pohl et W. Zyrnicki, Simultaneous determination of As, Bi, Se and Sn and non-hydride forming elements by HG-ICP-AES without liquid phase separation, CHEM ANAL, 45(5), 2000, pp. 699-708
A Jobin Yvon hydride generation system without a gas-liquid phase separator
was employed for the determination of As, Bi, Se and Sn by the ICP-AES met
hod under various experimental conditions. Influence of forward power, NaBH
4 flow-rate and sample flowrate on analytical line sensitivities and detect
ion limits of the determined elements was examined and discussed. Applicabi
lity of the JY IMAGE spectrum acquisition system to the determination of th
e hydride forming elements was investigated. The As, Bi, Se and Sn signal s
uppression due to matrix effects of Na, K, Mg and Ca was observed. The hydr
ide generator was used also for the simultaneous determination of hydride a
nd non-hydride forming elements.