J. Szpunar et al., VALIDATION OF THE DETERMINATION OF COPPER AND ZINC IN BLOOD-PLASMA AND URINE BY ICP MS WITH CROSS-FLOW AND DIRECT-INJECTION NEBULIZATION, Talanta, 44(8), 1997, pp. 1389-1396
The simultaneous determination of copper and zinc in human plasma and
urine by inductively coupled plasma mass spectrometry (ICP MS) is disc
ussed. The performances of a cross-flow nebulizer and a direct-injecti
on nebulizer (DIN) were compared. Flow-injection-DIN-ICP MS analysis o
f clinical samples using 1-2 mu l samples was optimized. Isobaric inte
rferences were discussed and were demonstrated to be eliminated for th
e Cu-65 and most of the Zn nuclides. The need for standard addition to
compensate for signal suppression in the case of some serum samples w
as indicated. Results obtained by ICP MS using calibration with aqueou
s standard solutions were found to be in good agreement with those obt
ained by flame AAS for a batch of real blood plasma and urine samples.
The methods developed were validated by analysis of several standard
reference materials. (C) 1997 Elsevier Science B.V.