A chemiluminescence (CL) flow system is described for the determination of
isoniazid based on its enhancement on the chemiluminescence (CL) emission p
roduced upon mixing a hexacyanoferrate(III) solution with an alkaline lumin
ol solution. The system responds linearly to isoniazid concentration in the
range 0-1 mg/L with a detection limit (3 sigma) of 0.03 mug/L, relative st
andard deviation (RSD) of 1.2% for 0.1 mg/L isoniazid (n = 11). The system
has been successfully applied to the determination of isoniazid in pharmace
utical preparations.