Design considerations for transmission x-ray tubes operated at diagnostic energies

Citation
Lmn. Tavora et al., Design considerations for transmission x-ray tubes operated at diagnostic energies, J PHYS D, 33(19), 2000, pp. 2497-2507
Citations number
32
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science
Journal title
JOURNAL OF PHYSICS D-APPLIED PHYSICS
ISSN journal
00223727 → ACNP
Volume
33
Issue
19
Year of publication
2000
Pages
2497 - 2507
Database
ISI
SICI code
0022-3727(20001007)33:19<2497:DCFTXT>2.0.ZU;2-H
Abstract
A low-energy expansion of the EGS4 Monte Carlo code system was used in stud ies that aimed to understand photon production in transmission target x-ray tubes. In each case, foil targets were modelled with incident electrons of a few tens of keV. The production of bremsstrahlung and fluorescence radia tion at different depths was studied and the angular distribution of radiat ion analysed. The characteristics of x-ray spectra measured along the beam path, for targets of different materials, thicknesses and several beam ener gies have been determined and the relation between efficiency and the fluor escence-to-bremsstrahlung ratio is shown.