Construction and analytical applications of plastic membrane electrodes for oxytetracycline hydrochloride

Citation
Ym. Issa et al., Construction and analytical applications of plastic membrane electrodes for oxytetracycline hydrochloride, MIKROCH ACT, 135(1-2), 2000, pp. 97-104
Citations number
18
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
MIKROCHIMICA ACTA
ISSN journal
00263672 → ACNP
Volume
135
Issue
1-2
Year of publication
2000
Pages
97 - 104
Database
ISI
SICI code
0026-3672(2000)135:1-2<97:CAAAOP>2.0.ZU;2-G
Abstract
Oxytetracycline hydrochloride-selective electrodes of both the coated wire and the conventional polymer membrane types based on oxytetracyclinium phos photungstate and phosphomolybdate have been prepared. A Nernstian response is shown by these electrodes within 1.0 x 10(-6) - 1.0 x 10(-2) M concentra tion ranges depending on the type of electrode. The response is unaffected by the change of pH over the range 4-11. The standard electrode potentials, E degrees, were determined at different temperatures and used to calculate the isothermal temperature coefficients of the electrodes, The electrodes show good selectivity to oxytetracycline hydrochloride with respect to many inorganic cations, sugars and amino-acids. Oxytetracycline hydrochloride i s determined successfully in pure solutions and in pharmaceutical preparati ons using calibration by standard addition and potentiometric titration. A regeneration process for the exhausted electrodes has been developed.