Pa. Levine et Mj. Moran, Device eccentricity: Postmagnet behavior of DDDR pacemakers with automaticthreshold tracking, PACE, 23(10), 2000, pp. 1570-1572
Citations number
1
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
The normal pacing system function associated with the AutoCapture Pacing Sy
stem algorithm triggered concern on the part of the clinical staff caring f
or the patient when the initiation of a threshold search sequence was detec
ted during a transtelephonic follow-up evaluation. The analysis of the rhyt
hm demonstrates that the behavior of the system is nor mal and consistent w
ith the design of the AutoCapture algorithm. As a variety of new algorithms
are introduced, similar unexpected but normal behaviors can be anticipated
.