Analysis of sub-boil distilled water, hydrochloric and nitric acid for trace element impurities by ICP-MS

Citation
V. Balaram et al., Analysis of sub-boil distilled water, hydrochloric and nitric acid for trace element impurities by ICP-MS, I J CHEM A, 39(5), 2000, pp. 567-570
Citations number
7
Categorie Soggetti
Chemistry
Journal title
INDIAN JOURNAL OF CHEMISTRY SECTION A-INORGANIC BIO-INORGANIC PHYSICAL THEORETICAL & ANALYTICAL CHEMISTRY
ISSN journal
03764710 → ACNP
Volume
39
Issue
5
Year of publication
2000
Pages
567 - 570
Database
ISI
SICI code
0376-4710(200005)39:5<567:AOSDWH>2.0.ZU;2-3
Abstract
A method has been designed for the precise estimation of a few trace and ul tra-trace elements in high purity sub-boil distilled water and hydrochloric and nitric acids using inductively coupled plasma mass spectrometers of tw o different makes. The trace element concentrations of majority of the elem ents analysed in sub-boil distilled water are found to be below detection l imits. A comparison of the trace element impurity levels obtained in hydroc hloric and nitric acids with the maximum allowable impurity levels for high purity reagents documented in SEMI, suggests that these purity levels conf irm with the specifications given for the high pure reagents.