Stem displacement during reduction of a dislocated cemented total hip arthroplasty with a polished tapered stem

Citation
Hm. Staal et al., Stem displacement during reduction of a dislocated cemented total hip arthroplasty with a polished tapered stem, J ARTHROPLA, 15(7), 2000, pp. 944-946
Citations number
7
Categorie Soggetti
Ortopedics, Rehabilitation & Sport Medicine
Journal title
JOURNAL OF ARTHROPLASTY
ISSN journal
08835403 → ACNP
Volume
15
Issue
7
Year of publication
2000
Pages
944 - 946
Database
ISI
SICI code
0883-5403(200010)15:7<944:SDDROA>2.0.ZU;2-3
Abstract
Displacement of the polished stem of a total hip arthroplasty from its ceme nted mantle occurred as a complication of an attempted closed reduction aft er luxation. No cement fractures were seen during subsequent reoperation. T he complication can be avoided if cement is placed over the shoulder of the prosthesis.