S. Yamato et al., On-line automated high-performance liquid chromatographic determination oftotal riboflavin phosphates using immobilized acid phosphatase as a pre-column reactor, J CHROMAT A, 896(1-2), 2000, pp. 171-181
An automated chromatographic detection system for the determination of tota
l riboflavin phosphates using immobilized sweet potato acid phosphatase as
a pre-column reactor is reported on. An immobilized enzyme reactor, incorpo
rated in the on-line analytical system, hydrolysed riboflavin phosphates to
riboflavin, and then lipophilic riboflavin was concentrated at the top of
an ODS trap column. Enzymatically hydrolysed riboflavin was back-eluted fro
m the trap column using a mobile phase containing methanol, and then subseq
uently chromatographed on an ODS analytical column. The effluents were moni
tored by UV absorption at 280 nm. The calibration graph for total riboflavi
n phosphates, determined by this method, was linear over the range 0.5-500
nmol/ml, with a correlation coefficient of 0.9999, The detection Limit at a
signal-to-noise ratio of 3 was 25 pmol/ml. The average conversion rate of
riboflavin phosphates to riboflavin was estimated at 97%. The relative stan
dard deviations of the intra- and inter-assay precision were 1.2 and 2.6%,
respectively. (C) 2000 Elsevier Science B.V. All rights reserved.