Evaluation of BactoScan FC for the enumeration of bacteria in raw milk

Citation
V. Ninane et al., Evaluation of BactoScan FC for the enumeration of bacteria in raw milk, LAIT, 80(5), 2000, pp. 527-538
Citations number
12
Categorie Soggetti
Food Science/Nutrition
Journal title
LAIT
ISSN journal
00237302 → ACNP
Volume
80
Issue
5
Year of publication
2000
Pages
527 - 538
Database
ISI
SICI code
0023-7302(200009/10)80:5<527:EOBFFT>2.0.ZU;2-Q
Abstract
The evaluation of BactoScan FC for routine enumeration of bacteria in raw m ilk was carried out according to the model proposed by the International Da iry Federation. The linearity of response of BactoScan FC was verified betw een 6 000 and 27 000 000 pulses.mL(-1) without reaching the limits of detec tion which have not been determined. The transmission error was estimated a t 0.09% and the repeatability at 0.084 log(10) CFU.mL(-1) for a 60 000 CFU. mL(-1) mean level of contamination. The stability of the equipment over tim e was verified; in both the short term and the long term, reproducibility i s in the same order of magnitude as repeatability. The relationship between BactoScan FC and the reference method is linear over the whole range analy sed (700 to 26 000 000 CFU.mL(-1)), with a determination threshold estimate d at under 4 000 CFU.mL(-1). The regression equation between the reference method and the BactoScan FC is: log(10) (CFU.mL(-1)) = 0.839 x log(10) (pul ses.mL(-1)) + 0.095. The accuracy of estimation of the reference value, exp ressed by the residual standard deviation of the regression, is 0.287 log(1 0) CFU.mL(-1). Finally, the independence of this estimate compared to the s omatic cell level of the samples was verified.