Mg. Mitch et al., Well-ionization chamber response relative to NIST air-kerma strength standard for prostate brachytherapy seeds, MED PHYS, 27(10), 2000, pp. 2293-2296
Citations number
4
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
The response of well-ionization chambers to the emissions of Pd-103 and I-1
25 radioactive seed sources used in prostate cancer brachytherapy has been
measured. Calibration factors relating chamber response (current or dial se
tting) to measured air-kerma strength have been determined for seeds from n
ine manufacturers, each with different designs. Variations in well-ionizati
on chamber response relative to measured air-kerma strength have been obser
ved because of differences in the emitted energy spectrum due to bath the r
adionuclide support material (I-125 seeds) and the mass ratio of Pd-103 to
Pd-102 (Pd-103 seeds). Obtaining accurate results from quality assurance me
asurements using well-ionization chambers at a therapy clinic requires know
ledge of such differences in chamber response as a function of seed design.
[S0094-2405(00)00910-X].