Mf. Moyers et al., Determination of the air w-value in proton beams using ionization chamberswith gas flow capability, MED PHYS, 27(10), 2000, pp. 2363-2368
Citations number
22
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
The purpose of this work was to determine the w-value of air for protons us
ing the paired gas method. Several plastic- and magnesium-walled chambers w
ere used with air, synthetic air, nitrogen, and argon flowing gases. Using
argon as a reference gas, the w-value of air was measured and ranged from 3
2.7 to 34.5 J/C for protons with energies encountered in radiotherapy. Usin
g nitrogen as a reference gas, the w-value of air ranged from 35.2 to 35.4
J/C over the same range of proton energies. The w-value was found, at a giv
en energy, to be independent of the ion chamber used. The uncertainty in th
ese measurements was estimated at 5.2% at the 2 sigma level. This uncertain
ty was dominated by the 4.4% uncertainty in the w-value of the reference ga
s. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)
01010-5].