Underestimation of equivalent dose in single-aliquot optical dating of feldspars caused by preheating

Citation
J. Wallinga et al., Underestimation of equivalent dose in single-aliquot optical dating of feldspars caused by preheating, RADIAT MEAS, 32(5-6), 2000, pp. 691-695
Citations number
15
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science
Journal title
RADIATION MEASUREMENTS
ISSN journal
13504487 → ACNP
Volume
32
Issue
5-6
Year of publication
2000
Pages
691 - 695
Database
ISI
SICI code
1350-4487(200010/12)32:5-6<691:UOEDIS>2.0.ZU;2-9
Abstract
The ability of single-aliquot techniques to recover a known Laboratory dose in coarse-grain potassium-rich feldspar separates is tested. It is shown t hat, for the samples under study, single-aliquot methods underestimate the known dose. It is deduced that this phenomenon is caused by an increase in electron trapping probability as a consequence of heating. The same phenome non would lead to an underestimation in the equivalent dose in a natural sa mple. Multiple-aliquot techniques for equivalent dose determination may not suffer from this problem, but are not suitable for samples that might not have been completely bleached at the time, of deposition. it is argued that testing for changes in electron trapping probability should become routine practice in the single-aliquot dating of feldspar. (C) 2000 Elsevier Scien ce Ltd. All rights reserved.