PURPOSE: To investigate how changes in kilovolt peak and milliampere second
settings, and patient weight affect transmitted x-ray energy fluence and t
he image contrast-to-noise ratio (CNR) at abdominal computed tomography (CT
).
MATERIALS AND METHODS: Cylinders of water were used as patient models, and
x-ray spectra, including x-ray tube potentials of 80-140 kVp, were investig
ated. The mean photon energy and energy fluence transmitted through water c
ylinders with varying diameters and the image contrast for fat, muscle, bon
e, and iodine relative to water were determined. The effect of changing the
x-ray tube potential on CNR also was investigated.
RESULTS: At a constant kVp, increasing patient weight from 10 kg to 120 kg
reduced the transmitted energy fluence by two orders of magnitude. Changing
the x-ray tube potential from 80 kVp to 140 kVp increased the mean photon
energy from approximately 52 keV to approximately 72 keV and thus reduced t
he image contrast relative to water by 12% for muscle, 21% for fat, 39% for
bone, and 50% for iodine (approximate reduction values). Increasing the x-
ray tube potential from 80 kVp to 140 kVp increased the CNR by a factor of
2.6 for muscle and by a factor of 1.4 for iodine.
CONCLUSION: With changes in patient weight at abdominal CT, x-ray tube pote
ntials must be varied to maintain a constant detector energy fluence. Incre
asing the x-ray tube potential generally improves CNR.