J. Seneviratne et al., Donnan dialysis preconcentration coupled with ion chromatography and electrocatalytic oxidation for the determination of cyanide, TALANTA, 52(6), 2000, pp. 1025-1031
An electrocatalytic amperometric detector for the ion chromatographic deter
mination of CN- is described. A conducting composite that is based on a gra
phite-loaded sol-gel material comprises the working electrode. The composit
e is doped with a Ru-II metallodendrimer which is demonstrated to promote t
he electrochemical oxidation of CN- at potentials positive of 0.5 V vs. Ag/
AgCl. In 6 mM NaOH, 0.05 M NaCl flowed at 1.0 ml min(-1), a 5-point calibra
tion curve with the following linear least squares parameters is obtained o
ver the range, 1.0-30 M CN-: slope, 24.2 +/- 0.1 nA M-1; intercept, - 6 +/-
2 nA; and r, 0.9997. The detection limit, 0.7 muM CN-, compares favorably
to that obtained by amperometry at a silver electrode, 0.5 muM CN-, under c
omparable experimental conditions. A 60-min preconcentration by Donnan dial
ysis increases the sensitivity by a factor of 23.6. (C) 2000 Elsevier Scien
ce B.V. All rights reserved.