Cs. Evans et al., Improved sensitivity in detection of chlormequat by liquid chromatography-mass spectrometry, J CHROMAT A, 897(1-2), 2000, pp. 399-404
Two published separations, using electrospray mass spectrometry (ES-MS), ex
hibit significant differences in Limits of detection (LODs) for chlormequat
cation in pear. Separation on ODS1, confirmed to result from ion-exchange,
gives shorter analysis times and calibration over a wider concentration ra
nge than on an SCX cation-exchange column. The superior LOD using ODS1 (0.0
4 ng ml(-1) vs. 1.0 ng ml(-1)) results mainly from better chromatographic p
eak shape. Separation on ODS1 combined with optimised ES-MS detection allow
s direct quantification of chlormequat on an ion trap instrument at levels
lower than those required for residue analysis in foods and also in drinkin
g water. (C) 2000 Elsevier Science B.V. All rights reserved.