Improved sensitivity in detection of chlormequat by liquid chromatography-mass spectrometry

Citation
Cs. Evans et al., Improved sensitivity in detection of chlormequat by liquid chromatography-mass spectrometry, J CHROMAT A, 897(1-2), 2000, pp. 399-404
Citations number
15
Categorie Soggetti
Chemistry & Analysis","Spectroscopy /Instrumentation/Analytical Sciences
Journal title
Volume
897
Issue
1-2
Year of publication
2000
Pages
399 - 404
Database
ISI
SICI code
Abstract
Two published separations, using electrospray mass spectrometry (ES-MS), ex hibit significant differences in Limits of detection (LODs) for chlormequat cation in pear. Separation on ODS1, confirmed to result from ion-exchange, gives shorter analysis times and calibration over a wider concentration ra nge than on an SCX cation-exchange column. The superior LOD using ODS1 (0.0 4 ng ml(-1) vs. 1.0 ng ml(-1)) results mainly from better chromatographic p eak shape. Separation on ODS1 combined with optimised ES-MS detection allow s direct quantification of chlormequat on an ion trap instrument at levels lower than those required for residue analysis in foods and also in drinkin g water. (C) 2000 Elsevier Science B.V. All rights reserved.