Simultaneous preconcentration and speciation of iron(II) and iron(III) in water samples by 2-mercaptobenzimidazole-silica gel sorbent and flow injection analysis system
H. Bagheri et al., Simultaneous preconcentration and speciation of iron(II) and iron(III) in water samples by 2-mercaptobenzimidazole-silica gel sorbent and flow injection analysis system, ANALYT CHIM, 424(2), 2000, pp. 233-242
2-Mercaptobenzimidazole, loaded on silica gel (MBI-SG) was used for preconc
entration and speciation of iron(II) and iron(III) in water samples. The sp
eciation is based upon selective retention of iron(III) on MBI-SG sorbent a
nd quantitatively passing iron(II) ions. This type of sorbent was not only
capable of separating theses two ions but also could concentrate them. The
adsorbed iron(III) on MBI-SG was desorbed from the column by an SCN- soluti
on and transferred to the detection device. The effluent, then, was treated
with hydrogen peroxide to oxidize iron(II) into iron(III) and subsequently
was pumped through the MBI-SG column. A home-made flow injection system wa
s coupled with a UV-VIS and an atomic absorption spectrometer, simultaneous
ly. This method was applied to four different water samples and sub-ppb det
ection limit, high accuracy (>99% recovery), good precision (R.S.D. < 5%),
and good potential for automation were observed. (C) 2000 Elsevier Science
B.V. All rights reserved.