Pr. Territo et Rs. Balaban, Rapid spectrophotometric determination of oxygen consumption using hemoglobin, in vitro: Light scatter correction and expanded dynamic range, ANALYT BIOC, 286(1), 2000, pp. 156-163
The method of using absorbance in conjunction with hemoglobin (Hb) to monit
or rapid changes in oxygen consumption in vitro was improved by using a non
linear calibration technique and multiwavelength spectroscopy. The O-2 depe
ndence of Hb absorbance was effectively linearized using the current techni
que (R-2 = 0.990 +/- 0.0021 n = 3), and extended the dynamic range of [O-2]
determinations by 1.6-fold over previous approaches. The association/disso
ciation rates of O-2 and Hb were evaluated using the current approach and w
ere not significant on the 100-ms time domain. A method was also developed
for compensating for large amplitude light scattering changes in turbid med
ia using multiwavelength analysis. Both the nonlinear calibration curve and
light scattering corrections were validated in isolated porcine heart mito
chondrial preparations. (C) 2000 Academic Press.