Jmt. Carneiro et al., Individual sample conditioning in flow analysis. Determination of N-total in plant materials, FRESEN J AN, 368(5), 2000, pp. 496-500
A flow-batch system allowing in-line individual sample matrix matching is p
roposed for analysis of sample lots with high variability in acidity. The f
easibility of the approach is demonstrated in the spectrophotometric determ
ination of total nitrogen in Kjeldahl digests, using a column with a slight
ly soluble reagent (AgCl). The solutions are sequentially injected by means
of an 8-port selecting valve and processed in a mixing chamber that is als
o used as a monitoring unit. The system yields reproducible results (r.s.d.
usually < 2.5%) and the sampling rate is 14 samples/h. The analytical curv
e is linear within 1.00 and 6.00% N (dry basis), and the regression coeffic
ient is > 0.999 (n = 6). Results are in agreement with certified values of
standard reference materials and with results obtained by conductometry.