The aims of intravascular ultrasound (IVUS) are visualization of the arteri
al wall, characterization of the arterial plaque, measurement of vessel dia
meter and endoluminal procedure assessment. The main sources of errors are:
limitation of resolution, distorsion of images due to non uniform angular
velocity of mechanical type transducers, overestimation or underestimation
of cross sectional area secondary to eccentric position or non coaxial orie
ntation of the catheter. Awareness of these problems may prevent misinterpr
etations.
Combined balloon imaging catheters, foward-looking sector scanners and ultr
asound imaging guide wires may be achieve by miniaturization of transducers
. Three dimensional imaging, blood dow quantification, local elasticity of
tissues may be achieve by new softwares.