Estimation of Tc-99m-MAG3 clearance by single-sample methods and camera-based methods
Citation
Y. Inoue et al., Estimation of Tc-99m-MAG3 clearance by single-sample methods and camera-based methods, ANN NUCL M, 14(5), 2000, pp. 329-332
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging
Journal title
ANNALS OF NUCLEAR MEDICINE
SICI code
0914-7187(200010)14:5<329:EOTCBS>2.0.ZU;2-Q
Abstract
We compared single-sample methods, proposed by Russell et al. and Bubeck ct
al., and camera-based methods in calculating Tc-99m-MAG3 clearance, and de
termined camera-based methods that provide estimates comparable to those me
asured by the Russell method. Twenty-one patients underwent Tc-99m-MAG3 ren
al scintigraphy, and clearance was measured by the Russell method and Bubec
k method. Various renogram parameters were determined based on the slept: o
f the renogram and area under the renogram, and correlated with the clearan
ce measured by the Russell method. Camera-based clearance was calculated wi
th the obtained regression equations and with equations determined previous
ly using the Bubeck method as a standard. The Bubeck method provided lower
measures than the Russell method in high renal function. Clearance measured
by die Russell method was well correlated with renogram parameters, and cl
earance calculated with the obtained regression equation was comparable to
that measured by the Russell method. When camera-based clearance was predic
ted with the previous equation, it was lower than the result obtained by th
e Russell method in high function. In conclusion, there are systematic diff
erences in Tc-99m-MAG3 clearance calculated by different methods. The camer
a-based methods obtained in this study appear to facilitate: comparison of
results obtained by the Russell method and camera-based method.