A train of carbon and DNPH-coated cartridges for the determination of carbonyls from C-1 to C-12 in air and emission samples

Citation
M. Possanzini et al., A train of carbon and DNPH-coated cartridges for the determination of carbonyls from C-1 to C-12 in air and emission samples, ATMOS ENVIR, 34(29-30), 2000, pp. 5311-5318
Citations number
27
Categorie Soggetti
Environment/Ecology,"Earth Sciences
Journal title
ATMOSPHERIC ENVIRONMENT
ISSN journal
13522310 → ACNP
Volume
34
Issue
29-30
Year of publication
2000
Pages
5311 - 5318
Database
ISI
SICI code
1352-2310(2000)34:29-30<5311:ATOCAD>2.0.ZU;2-N
Abstract
A train of carbon and DNPH-coated cartridges set in series is proposed as a suitable method for determining all carbonyl compounds from C-1 to C-12 in air and emission samples. While the front traps are analyzed by GC-MS afte r thermal desorption, back-up cartridges are analyzed by HPLC-UV after liqu id extraction. The method has been tested through laboratory and field expe riments. Practical applications are reported and data obtained used for dis cussing the possible origin of volatile and semi-volatile compounds in fore st areas of the Mediterranean basin. (C) 2000 Elsevier Science Ltd. All rig hts reserved.