Specifications for equipment used for infant pulmonary function testing

Citation
U. Frey et al., Specifications for equipment used for infant pulmonary function testing, EUR RESP J, 16(4), 2000, pp. 731-740
Citations number
22
Categorie Soggetti
Cardiovascular & Respiratory Systems","da verificare
Journal title
EUROPEAN RESPIRATORY JOURNAL
ISSN journal
09031936 → ACNP
Volume
16
Issue
4
Year of publication
2000
Pages
731 - 740
Database
ISI
SICI code
0903-1936(200010)16:4<731:SFEUFI>2.0.ZU;2-J
Abstract
The aim of this position paper is to define minimal performance criteria fo r the separate items comprising equipment used to measure respiratory funct ion in infants together,vith overall performance criteria for the assembled pieces of such equipment. These guidelines cover numerous aspects including: 1) safety, 2) documentat ion and maintenance of equipment, 3) physical characteristics of mechanical parts and signal transducers, and 4) data acquisition, Further, validation procedures for individual components as well as for the integrated equipme nt are recommended, Adherence to these guidelines should ensure that infant lung function measu rements can be performed with an acceptable degree of safety, precision and reproducibility. They will also facilitate multicentre collection of data and performance of clinical investigations. Manufacturers of infant respiratory function equipment should make every ef fort to comply with these guidelines, which represent the current standards of paediatric health professionals in this field.