A mass spectrometer using an ITD cell was constructed, and the mass spectru
m of Ca+ ions was measured using the mass selective instability scan method
. The distance between the two end-cap electrodes and the diameter of the r
ing electrode are 14.14 mm (=2z(0)) and 20 mm (=2r(0)), respectively. The f
requency of the RF(radio frequency) voltage applied to the ring electrode i
s fixed at 1.1 MHz. The optimum value of the He buffer gas pressure for the
sensitivity and resolution was experimentally determined to be 3.0 x 10(-4
) Torr. The possibility of substituting air for the He buffer gas was exami
ned.