NPL-CSIC comparison of regular reflectance measurements

Citation
J. Campos et al., NPL-CSIC comparison of regular reflectance measurements, METROLOGIA, 37(4), 2000, pp. 323-327
Citations number
3
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
METROLOGIA
ISSN journal
00261394 → ACNP
Volume
37
Issue
4
Year of publication
2000
Pages
323 - 327
Database
ISI
SICI code
0026-1394(2000)37:4<323:NCORRM>2.0.ZU;2-D
Abstract
A comparison of two different scales of regular reflectance measurements ha s been made between the National Physical Laboratory (NPL, UK) and the Cons ejo Superior de Investigaciones Cientificas (CSIC, Spain). Both laboratorie s have spectrally calibrated a reflectance transfer standard at angles of i ncidence of 8 degrees and 45 degrees in the wavelength range 350 nm to 1000 nm. Relative differences in reflectance are between 2 parts in 10(3) and - 2 parts in 10(3) at 8 degrees, and between 1 part in 10(3) and -4 parts in 10(3) at 45 degrees. These results agree well within their respective uncer tainties.