A 7-channel magnetometer for magnetocardiography based on high-T-c SQU
IDs has been realized, This magnetometer is used for test experiments
in the development of a multichannel high-T-c SQUID based heart-scanne
r for clinical applications, The intrinsic noise level of the channels
in the 7-channel system is typically 120 fT/root Hz down to 1 Hz. Mag
netocardiograms were recorded inside a magnetically shielded room, Int
roductory experiments were performed on the suppression of noise by co
mbining magnetometers to form planar gradiometers. The noise suppressi
on that can be established appeared to be limited by the imbalance of
the gradiometric configuration, which is roughly 2%. This relatively p
oor balance of the system is caused by inaccuracies in the transfer fu
nctions of the individual SQUID magnetometers, and by deviations from
the planar geometry.