A new beam line and injector deck for AMS measurements have been built at t
he 8 MV tandem accelerator of the National Institute of Nuclear Physics and
Engineering, Bucharest, Romania. The main components on the low-energy sid
e are a high-current cesium sputter source, a 90 degrees injection magnet a
nd a pre-acceleration stage. At the high-energy side the beam line is achro
matic, consisting of two 90 degrees analysing magnets with mass energy prod
uct 120 MeV amu and a gas-filled ionization chamber. The system will be com
plete with a Wien filter and a multi-anode gas detector with time-of-flight
discrimination. Presently, the AMS facility is undergoing tests and routin
e measurements are expected to start soon. (C) 2000 Elsevier Science B.V. A
ll rights reserved.