Detection of heavy elements by accelerator mass spectrometry with the elect
ron cyclotron resonance ion source, Argonne linear accelerator and fragment
mass analyzer (ECRIS-ATLAS-FMA) system has been developed. The use of the
ECR-ATLAS system for AMS of heavy elements has two interesting features: (i
) the efficient production of high-charge state ions in the ECR source ensu
res the elimination of molecular ions at the source stage, a highly attract
ive feature for any mass-spectrometric use not exploited so far; (ii) the l
inear acceleration based on velocity matching and the beam transport system
act as a powerful mass filter for background suppression. We have shown th
at our system reaches an abundance sensitivity of 1 x 10(-14) for Pb isotop
es. The U-236 detection sensitivity is U-236/U greater than or similar to 1
x 10(-12), limited mainly by the ion source output. (C) 2000 Elsevier Scie
nce B.V. All rights reserved.