Mj. Butson et al., Verification of lung dose in an anthropomorphic phantom calculated by the collapsed cone convolution method, PHYS MED BI, 45(11), 2000, pp. N143-N149
Verification of calculated lung dose in an anthropomorphic phantom is perfo
rmed using two dosimetry media. Dosimetry is complicated by factors such as
variations in density at slice interfaces and appropriate position on CT s
canning slice to accommodate these factors. Dose in lung for a 6 MV and 10
MV anterior-posterior field was calculated with a collapsed cone convolutio
n method using an ADAC Pinnacle, 3D planning system. Up to 5% variations be
tween doses calculated at the centre and near the edge of the 2 cm phantom
slice positioned at the beam central axis were seen, due to the composition
of each phantom slice. Validation of dose was performed with LiF thermolum
inescent dosimeters (TLDs) and X-Omat V radiographic film. Both dosimetry m
edia produced dose results which agreed closely with calculated results nea
rest their physical positioning in the phantom. The collapsed cone convolut
ion method accurately calculates dose within inhomogeneous lung regions at
6 MV and 10 MV x-ray energy.