L. Ding et al., Comparison of a low microwave-powered thermospray nebulizer and a pneumatic nebulizer for ICP-AES, CHEM J CH U, 21(11), 2000, pp. 1661-1663
The analytical performance of two sample introduction systems, a low-microw
ave-powered thermospray nebulizer and a pneumatic nebulizer, coupled to an
inductively coupled plasma atomic emission spectrometer, was compared. Seve
ral parameters including the sample uptake rate, the carrier gas flow rate
and the observation height:(above the load coil) were optimized and compare
d. Under the optimized conditions, the background equivalent concentration
of Mn 257.610 nm, memory effect, HCl concentration effect on emission inten
sity of some selected elements, detection limits (DLs) and relative standar
d deviation (RSD) for Cd, Co, Cr, Mn, La with either nebulizer were studied
. The DLs obtained with MWTN are lower than those with PN, but the RSDs are
worse with MWTN than with PN.