Comparison of a low microwave-powered thermospray nebulizer and a pneumatic nebulizer for ICP-AES

Citation
L. Ding et al., Comparison of a low microwave-powered thermospray nebulizer and a pneumatic nebulizer for ICP-AES, CHEM J CH U, 21(11), 2000, pp. 1661-1663
Citations number
9
Categorie Soggetti
Chemistry
Journal title
CHEMICAL JOURNAL OF CHINESE UNIVERSITIES-CHINESE
ISSN journal
02510790 → ACNP
Volume
21
Issue
11
Year of publication
2000
Pages
1661 - 1663
Database
ISI
SICI code
0251-0790(200011)21:11<1661:COALMT>2.0.ZU;2-0
Abstract
The analytical performance of two sample introduction systems, a low-microw ave-powered thermospray nebulizer and a pneumatic nebulizer, coupled to an inductively coupled plasma atomic emission spectrometer, was compared. Seve ral parameters including the sample uptake rate, the carrier gas flow rate and the observation height:(above the load coil) were optimized and compare d. Under the optimized conditions, the background equivalent concentration of Mn 257.610 nm, memory effect, HCl concentration effect on emission inten sity of some selected elements, detection limits (DLs) and relative standar d deviation (RSD) for Cd, Co, Cr, Mn, La with either nebulizer were studied . The DLs obtained with MWTN are lower than those with PN, but the RSDs are worse with MWTN than with PN.