High-accuracy primary and transfer standards for radiometric calibration
Citation
Xb. Zheng et al., High-accuracy primary and transfer standards for radiometric calibration, CHIN SCI B, 45(21), 2000, pp. 2009-2013
Categorie Soggetti
Multidisciplinary
Journal title
CHINESE SCIENCE BULLETIN
SICI code
1001-6538(200011)45:21<2009:HPATSF>2.0.ZU;2-A
Abstract
The first research and experimental results obtained in China of high-accur
acy radiometric calibration based on cryogenic radiometer are reported. Unc
ertainties of cryogenic radiometer and trap detectors at 7 wavelengths in t
he visible spectrum (488-786 nm) were less than 0.023% and 0.035% respectiv
ely which proved the reasonability and possibility of establishing and tran
sferring high-accuracy radiometric standards based on detectors.