Rq. Wang et Xa. Li, A Monte Carlo calculation of dosimetric parameters of Sr-90/Y-90 and Ir-192 SS sources for intravascular brachytherapy, MED PHYS, 27(11), 2000, pp. 2528-2535
Citations number
18
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
The AAPM TG-60 report has proposed various dose calculation parameters for
intravascular brachytherapy (IVBT). These parameters include the dose rate
constant (or the dose rate at a reference position for a beta-particle emit
ting source), the radial dose function, and the anisotropy function. In thi
s work, we have used a modified EGS4 Monte Carlo system to calculate these
parameters for the two most commonly used IVBT sources (the beta-particle e
mitting Sr-90/Y-90 source and the photon emitting Ir-192 SS source). To ens
ure the calculation accuracy, the present calculation was compared with sev
eral measurements and calculations reported by other authors. Excellent agr
eement was found for the results with the photon source. For beta-particle
source calculation, the present results for a variety of point sources agre
e very well with a previous work. The presently calculated radial dose func
tions for the Sr-90/Y-90 source are consistent with those of a published wo
rk for intermediate radial distances. The dose uniformity in the axial dire
ction was also studied. The contributions of bremsstrahlung photons to tota
l doses for the Sr-90/Y-90 beta source, and the influence of ignoring elect
ron transport on calculated doses for the Ir-192 SS photon source are discu
ssed. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(
00)01211-6].