Purpose. To assess use of the tonometer ProTon for measuring intraocular pr
essure in rabbits.
Materials and methods. Eleven chinchilla-bastard rabbits were measured unde
r standardized conditions with the applanation tonometer ProTon for 62+/-15
days. One rabbit was excluded because of intensively defensive reactions.
Results. Intraocular pressure in the ten rabbits was 12.15+/-3.63 mmHg in t
he right eye and 12.11+/-3.52 mmHg in the left eye.
Conclusion. Intraocular pressure can be measured reliably over a long perio
d with the applanation tonometer ProTon. It is therefore suitable for deter
mining intraocular pressure in experimentally induced glaucoma in rabbits.