Origin of a signal intensity loss artifact in fat-saturation MR imaging

Citation
L. Axel et al., Origin of a signal intensity loss artifact in fat-saturation MR imaging, RADIOLOGY, 217(3), 2000, pp. 911-915
Citations number
10
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Journal title
RADIOLOGY
ISSN journal
00338419 → ACNP
Volume
217
Issue
3
Year of publication
2000
Pages
911 - 915
Database
ISI
SICI code
0033-8419(200012)217:3<911:OOASIL>2.0.ZU;2-U
Abstract
Artifactual water signal intensity loss can be observed on fat-saturation m agnetic resonance (MR) images of inhomogeneous regions such as the thorax. Magnetic effects of air inclusions on fat-saturation pulses were investigat ed as the possible origin of this artifact. Computer simulation results agr eed well with observed production of water saturation by means of nominal f at suppression in MR imaging of phantoms and a representative clinical exam ple.