Cetylpyridinium-iodomercurate PVC membrane ion selective electrode for thedetermination of cetylpyridinium cation in Ezafluor mouth wash and as a detector for some potentiometric titrations
Mn. Abbas et al., Cetylpyridinium-iodomercurate PVC membrane ion selective electrode for thedetermination of cetylpyridinium cation in Ezafluor mouth wash and as a detector for some potentiometric titrations, TALANTA, 53(2), 2000, pp. 425-432
A potentiometric method is described for the determination of cetylpyridini
um (CP) cation using a polyvinyl chloride powder (PVC) membrane sensor base
d on CP-iodomercurate ion pair as an electroactive material. The sensor sho
ws a linear response for CP ion over the concentration range of 2 x 10(-6)-
1 x 10(-3) M at 25 degreesC. The electrode posses a sub-Nernstian slope of
29.0 +/- 0.4 mV decade(-1), which has been thoroughly discussed. The electr
ode shows a fast potential response of similar to 50 s, which is almost con
stant over a pH range of 3.0-6.0. Selectivity coefficient data for some com
mon ions show negligible interference, however, cetyltrimethylammonium brom
ide (CTMAB) interferes significantly. An average recovery of 98.8% for CP w
ith an average relative standard deviation (R.S.D.) of 1.2% has burn achiev
ed. The determination of CPC in Ezafluor mouth wash gave result that compar
e favorably with those obtained by the British Pharmacopoeia method. The CP
electrode has been utilized as an end point indicator electrode in potenti
ometric titration of some anions, and applied for the determination of anio
nic surfactants in some commercial detergents and waste water. (C) 2000 Els
evier Science B.V. All rights reserved.