M. Maizels et al., Spectroelectrochemical sensing based on multimode selectivity simultaneously achievable in a single device. 7. Sensing of Fe(CN)(6)(4-), ELECTROANAL, 12(17), 2000, pp. 1356-1362
A spectroelectrochemical sensor that has trimodal selectivity (selective pa
rtitioning, electrochemistry, and spectroscopy) is evaluated for the determ
ination of ferrocyanide in solution. The sensor is based on attenuated tota
l reflection spectroscopy at an indium-tin oxide optically transparent elec
trode coated with a cationic PDMDAAC-SiO2, where PDMDAAC means poly(dimethy
ldiallylammonium chloride) film into which anionic Fe(CN)(6)(4-) partitions
. Fe(CN)(6)(4-) loaded into the film is subjected to spectroelectrochemical
modulation, and the absorbance change at 320 nm associated with cycling Fe
(CN)(6)(4-)/Fe(CN)(6)(3-) is used to quantitate the analyte. Cyclic voltamm
ograms of Fe(CN)(6)(4-), 0.1 M KNO3 at high concentrations showed multiple
peaks, which are interpreted as two types of Fe(CN)(6)(4-)/Fe(CN)(6)(3-) in
the PDMDAAC-SiO2 film: Fe(CN)(6)(4-) bound to PDMDAAC and free Fe(CN)(6)(4
)- in interstitial regions of the porous film. The sensor response was affe
cted by ionic strength and anion type of the supporting electrolyte and fil
m thickness. The diffusion coefficient of Fe(CN)(6)(4-) in the PDMDAAC-SiO2
film was determined to be 4.9 x 10(-11) cm(2)/s. An analytical calibration
plot of absorbance change at 420 nm versus Fe(CN)(6)(4-) concentration wit
h a preconcentration time to equilibration was linear from 5 to 400 muM fer
rocyanide with negative deviation from linearity observed at higher concent
rations. However, a linear calibration plot at high concentrations (1-8 mM)
could be achieved by making sensor measurements in a very short time perio
d, only 1 min after exposure to the sample.