M. O'Brien et P. Lamperti, Comparison of the NIST and NPL air kerma standards used for X-ray measurements between 10 kV and 80 kV, J RES NAT I, 105(5), 2000, pp. 701-707
A direct comparison was made between the air kerma primary standards used f
or the measurements of low-energy x rays at the National Institute of Stand
ards and Technology (NIST) and the National Physical Laboratory (NPL). The
comparison was conducted at the NPL using NPL reference radiation qualities
between 10 kV and 80 kV. The results show the primary air-kerma standards
to agree within 0.6 % of their values for beam qualities up to 80 kV.