Comparison of the NIST and NPL air kerma standards used for X-ray measurements between 10 kV and 80 kV

Citation
M. O'Brien et P. Lamperti, Comparison of the NIST and NPL air kerma standards used for X-ray measurements between 10 kV and 80 kV, J RES NAT I, 105(5), 2000, pp. 701-707
Citations number
6
Categorie Soggetti
Multidisciplinary,"Engineering Management /General
Journal title
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY
ISSN journal
1044677X → ACNP
Volume
105
Issue
5
Year of publication
2000
Pages
701 - 707
Database
ISI
SICI code
1044-677X(200009/10)105:5<701:COTNAN>2.0.ZU;2-U
Abstract
A direct comparison was made between the air kerma primary standards used f or the measurements of low-energy x rays at the National Institute of Stand ards and Technology (NIST) and the National Physical Laboratory (NPL). The comparison was conducted at the NPL using NPL reference radiation qualities between 10 kV and 80 kV. The results show the primary air-kerma standards to agree within 0.6 % of their values for beam qualities up to 80 kV.